AirLife
Passive humidification
Bypass heat and moisture exchangers
Air Life Bypass HME Air Life Bypass HME (BHME) technology delivers aerosolized medication treatments without breaking the ventilator circuit. Easy one-handed operation allows bypass engagement for simple medication delivery set-up. The bypass lever position and clear visual indicators help identify gate engagement. The filtered BHME helps minimize the risk of respiratory complications and reduces the risk of cross-contamination.
Cat. no. 003020 003021
Description
Qty.
Air Life nonfiltered bypass HME Air Life filtered bypass HME
25/cs 25/cs
Air Life BHME humidifiers
Aerosol mode
Gate open
Aerosol mode
Lever down
HME mode
Gate closed
HME mode
Optional filter helps minimize the risk of respiratory complications and reduces chance of cross-contamination.
Lever up
Bypass lever position and clear visual indicators ensure gate engagement.
003020 Air Life Nonfiltered BHME
003021 Air Life Filtered BHME
003020 Air Life Nonfiltered BHME
003021 Air Life Filtered BHME
Tidal volumes
250–1,250 mL
250–1,250 mL
250 mL = 9.1 mg H 2 250 mL = 9.1 mg H 2 250 mL = 9.1 mg H 2
O/L O/L O/L
Filtration efficiency
N/A
Weight
43 g
44 g
Recommended change out
Dead space
68 mL
68 mL
24 hours
24 hours
250 mL = 9.1 mg H 2 O/L 500 mL = 10.6 mg H 2 O/L 750 mL = 12.9 mg H 2 O/L 1,000 mL = 14.3 mg H 2 O/L 1,250 mL = 14.9 mg H 2 O/L 250 mL = 35.0 mg H 2 500 mL = 33.5 mg H 2 750 mL = 31.2 mg H 2 O/L 1,000 mL = 29.8 mg H 2 O/L 1,250 mL = 29.2 mg H 2 O/L O/L O/L
250 mL = 8.7 mg H 2 O/L 500 mL = 11.1 mg H 2 O/L 750 mL = 12.9 mg H 2 O/L 1,000 mL = 14.7 mg H 2 O/L 1,250 mL = 15.7 mg H 2 O/L 250 mL = 35.4 mg H 2 500 mL = 33.0 mg H 2 750 mL = 31.2 mg H 2 O/L 1,000 mL = 29.4 mg H 2 O/L 1,250 mL = 28.4 mg H 2 O/L O/L O/L
15 mm ISO male connector for Y piece 15/22 mm ISO female/male connector for ET tube, closed suctioning system, etc.
Connections
Moisture loss
Gas leakage
0 cc/min at 7 kPa (71 cmH 2 O)
0 cc/min at 7 kPa (71 cmH 2 O)
Compliance
0.68 mL/kPa at 7 kPa (71 cmH 2 O)
0.68 mL/kPa at 7 kPa (71 cmH 2 O)
Moisture output
Field strength = 3 – Tesla or less Spatial gradient = 720 Gauss/cm or less Radio frequency (RF) field = 128 MHz Specific absorption rate = 2.9 W/kg
MR conditional
Flow at 0.5 L/sec = 0.6 cmH 2 O Flow at 1.0 L/sec = 1.6 cmH 2 O Flow at 1.5 L/sec = 3.3 cmH 2 O
Flow at 0.5 L/sec = 1.3 cmH 2 O Flow at 1.0 L/sec = 3.3 cmH 2 O Flow at 1.5 L/sec = 6.1 cmH 2 O
Resistance before use
Packaging
25 units per case
25 units per case
Flow at 0.5 L/sec = 1.5 cmH 2 O Flow at 1.0 L/sec = 4.0 cmH 2 O Flow at 1.5 L/sec = 6.0 cmH 2 O
Flow at 0.5 L/sec = 0.7 cmH 2 O Flow at 1.0 L/sec = 2.0 cmH 2 O Flow at 1.5 L/sec = 4.5 cmH 2 O
Resistance after use